  Data showing increased sensitivity

Data courtesy of: Harvard Medical School, Partners Healthcare System
Cambridge, MA
 
     
 
   
Coverslip Hybridization @ 42° C, 16hrs.   MAUI Hybridization @ 42° C, 16hrs.
 
Experimental results

1. Average spot size detection increase = 15%

2. Average raw signal Cy5 increase = >5X

3. Average raw signal Cy3 increase = >5X

4. Average background = same

5. Spot detection increase = 2.3X (from 32% to 73%)
 
     
 
 
 

Data showing reduced hybridization times

Data courtesy of: DNA Chip Research, Inc., JAPAN.

Coverslip Hybridization
16hrs./Cy3/Scanner PMT65
MAUI® Hybridization System
4hrs./Cy3/Scanner PMT65
 
 
 
 

Data showing increased sensitivity due to mixing

Data Courtesy of: The Analytical Chemistry article, Gains in sensitivity with a device that mixes microarray hybridization solution in a 25-um-thick chamber. Anal. Chem. 74: 6413-6417.

 
     
  Mixing of the hybridization solution significantly improves sensitivity from low abundance labeled targets. The following quantity of labeled bacteriophage lambda PCR product targets were used (fMol/slide): A = 22, B = 1.1, C = 0.10, D = 0.0094. Every microarray spot contained approximately 0.1 fMol (100 aMol) probe. Error bar represents one standard deviation of 22 identical microarray spots.  
     
     
 